JOAN LAB Touch System
JOAN LAB Touch System
FA2004T
Electronic Analytical Balance
USER MANUAL
ⅠAbout the Balance
Specification
| Model | Range | Division | Repeatability | Linearity | Pan Size |
| FA1104 | 110g | 0.1mg | 0.2mg | 0.5mg | ø 80mm |
| FA1204 | 120g | 0.1mg | 0.2mg | 0.5mg | ø 80mm |
| FA2004 | 200g | 0.1mg | 0.2mg | 0.5mg | ø 80mm |
| FA2104 | 210g | 0.1mg | 0.2mg | 0.5mg | ø 80mm |
| FA2204 | 220g | 0.1mg | 0.2mg | 0.5mg | ø 80mm |
Stainless Steel Pan
Super 5 inch Touch panel display
RS232 Interface
Mains adapter supplied as standard
Windshield supplied as standard
Height adjustable feet
Internal Auto Calibration
Selectable measure units:mg, g, oz, ct
Memory for accumulated time
Applications
Weighing
Net weight / tare
Below balance weighing
Piece counting function
Density Test
Percentage Test
9.1 DATA OUTPUT
| 1 | Model or a decimal point |
| 2 | A space or a decimal point |
| 3 | A space or * |
| 4 | + or - or a decimal point |
| 5 | data |
| 6 | Data or a decimal point |
| 7 | Data or a decimal point |
| 8 | Data or a decimal point |
| 9 | Data or a decimal point |
| 10 | Data or a decimal point |
| 11 | Data or a decimal point |
| 12 | Data |
| 13 | Unit 1 |
| 14 | Unit 2 |
| 15 | Unit 3 |
| 16 | Enter |
| 17 | Wrap |
We can choose the suitable baud rate and communication style for
different print or connect with the computer
8.1 FAULT JUDGMENT
If you come across any problem, you can check it by you self and find the
reasons.
| FAULT | RESON | EXCLUDE |
| No display | Not connected to the power supply; Fuse is broken; Power transformer damage; | Plug in the power line; Replace the fuse; Replacement of power transformer; Contact factory |
| Weighing unstable | Bad working conditions; The wind screen is open; Something between the table and balance; The power unstable; Weighing unstable; | Keep the environment stable,close the windows and doors; Close the glass door; Take away the things; Connect the stable power; |
| The weighing digitsis wrong | The balance not calibration. Not tare before weighing. No adjust the level. | Calibration. Tare before weighing. Adjust the level feet. |
ⅡSETTINGS
TABLE OF CONTENTS
1.0 KNOW YOUR BALANCE
2.0 SPECIFICATIONS
3.0 UNPACKING THE BALANCE
4.0 CALIBRATION THE BALANCE
5.0 LOCATING THE BALANCE
6.0 SETTING UP THE BALANCE
6.1 ASSEMBLING THE BALANCE
6.2 LEVELLING THE BALANCE
7.0 OPERATION SETTINGS
7.1 SENSITIVITY SETTING
7.2 SPEED SETTING
7.3 WEIGHING UNIT SETTING
7.4 COUNTING FUNCTION
7.5 PERCENTAGE FUNCTION
7.6 DENSITY FUNCTION
8.1 FAULT JUDGMENT
9.1 DATA OUTPUT
1.0 KNOW YOUR BALANCE
Thank you for selecting the FA-T Series Analytical Balance.
This Instruction Manual will guide you of the installation, accessories, trouble-shooting, after sales service information, general maintenance of the balance, etc. it will also guide you through the various applications.
Please read this Manual thoroughly before starting the operations. If you need any clarifications, feel free to contact us.
The FA-T balances are ideal for laboratory and general purpose weighing. The balances can also be used for some advanced weighing functions.
FEATURES:
Stainless Steel Pan
Super bright LCD display with backlight
RS232 Interface
Adapter supplied as standard
Windshield supplied as standard
Height adjustable feet, Bubble Level
Internal Auto Calibration
Selectable measure units, g, oz, ct,mg
Memory for accumulated time
Weighing fuction
Piece counting function
Below balance weighing
7. 5 PECENTAGE FUNCTION
Check the weight percentage function:
1.Choose the 100% which you will put on the pan
2.Touch the weighing value you want weighing,here is the sample weight,after the weighing stable touch again, the result will check and store.
3.Touch here and weighing the total weight, and touch again check the weight.
4.The percentage will be show here.
7. 6 Density Function.
Same operation with counting function,just touch the air weight and weighing in the air, then weighing in the liquid, the result will show in the display.
7.2 SPEED SETTING
---Highest Speed --Normal Speed --Low Speed
7.3 WEIGHING UNIT SETTING
Under the weighing state, press “CONVERSION” to choose the unit.
There are 4 units :g/ct/oz/mg.
7. 4 COUNTING FUNCTION
There are counting functions in the balance, And please keep the products
have the same weight, and the minimum weight must be ≥0.5mg.
1.Choose the sample quantity which you will put on the pan
2.Touch the weighing value you want weighing,here is the sample weight,after
the weighing stable touch again, the result will check and store.
3.Touch here and weighing the total weight, and touch again check the weight.
4.The quantity will be show here.
2.0 SPECIFICATIONS
| Model | FA1004 | FA1104 | FA1204 | FA2004 | FA2104 | FA2204 |
| Range | 0-100g | 0-110g | 0-120g | 0-200g | 0-210g | 0-220g |
| Division | 0.1mg | |||||
| Repeatability | 0.2mg | |||||
| Linearity | 0.5mg | |||||
| Resp. Time | ≤ 4 Sec. | |||||
| Pan Size | ø80mm | |||||
| Interface | RS-232 bi-directional | |||||
| Overall Dimensions | 340 x 215 x 350mm | |||||
| Operating temperature | +10°C to 40°C | |||||
| Power Supply | Output voltage 5 VDC, 600mA through External Power Adapter as standard (Input Voltage220 V) | |||||
| Net Weight | 8.5 kg | |||||
Remove the balance from the packing by carefully lifting it out of the box.
Inside the box you will find everything needed to start using the balance
4.0 CALIBRATION THE BALANCE
Press”CAL” key, and it will show and shining: CAL-XXX.XXXX,and the balance begin internal calibration.
You also can choose manual calibration with standard weight, and there will have the notes tell you step by step:
5.0 LOCATING THE BALANCE
6.0 SETTING UP THE BALANCE
6.1 ASSEMBLING THE BALANCE
6.2 LEVELLING THE BALANCE
After placing the balance in a suitable place, level it by using the bubble level. To level the balance turn the two adjustable feet at the rear of the balance until the bubble in the bubble level is center.
7.0 OPERATION SETTINGS
1.State Icon 2. Function State 3.Weighing Digits
4.Notes 5.Operation Keys
7.1 SENSITIVITY SETTING
--Highest Sensitivity -Normal Sensitivity -Sensitivity